Molecular Devices™ QPix 450 and QPix 460 Microbial Colony Picking System
You rely on your equipment. Your equipment should rely on ATG.
At Analytical Technologies Group, we specialize in maintenance and repair of the QPix 450 and QPix 460 Microbial Colony Picking System from Molecular Devices. By concentrating on this high-precision platform and keeping a comprehensive stock of spare parts, we can provide fast, reliable service that keeps your lab running smoothly.
Precision instruments like the QPix 450 and 460 demand routine, expert attention to maintain accurate colony detection and imaging. Even small issues can bring workflows to a standstill. Thousands of researchers rely on ATG to ensure to keep their instruments running seamlessly. You should too.
Tips for keeping your Molecular Devices™ QPix 450 and QPix 460 Microbial Colony Picking System in good working order:
Daily Maintenance:
Ensure that the interior of the instrument is free from dirt and dust. Check the surface of the x-drive drag-chain support bracket. Dust and debris collected here can be swept onto the instrument bed during operation.
Before storing or cleaning the head, wipe the exterior of the head with a cloth using 70% ethanol. To store the head, slide it into its metal cover with the pins facing inward and then secure it in place with the thumbscrew and washer.
Remove, clean, and sanitize the wash bath and brushes.
Weekly Maintenance:
If you use the external air compressor, drain the accumulated moisture from the internal air tank weekly.
Semi-Annual Maintenance:
Completely disassemble the head and thoroughly clean all component parts.
Having technical problems with your Molecular Devices™ QPix 450 and QPix 460 Microbial Colony Picking System?
Fast Response
We provide very prompt responses to customers within one business day.
Fast Turnaround
Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.
Expert Staff
ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.
Parts In Stock
To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.
Molecular Devices™ QPix 450 and QPix 460 Microbial Colony Picking System Troubleshooting Tips:
If one or more of the axes do not move:
Make sure that the main power switch is turned on at the wall, the Emergency Stop button is pulled out, and the Start button is pressed.
If the pins bend when inoculating in plate wells:
Make sure that the plate is aligned correctly towards both arrows in the plate holder.
Make sure that the correct plate, head, and spacer blocks are installed.
In case of low air warning:
If you are using a laboratory built-in air system, verify that the supplied air pressure is from 0.69 MPa to 0.83 MPa (100 psi to 120 psi) and that the control valve is switched on.
If you are using a standalone air compressor, verify that the compressor is switched on and that the supplied air pressure is from 0.69 MPa to 0.83 MPa (100 psi to 120 psi).