Molecular Devices™ QPix 420 Microbial Colony Picking System

You rely on your equipment. Your equipment should rely on ATG.

At Analytical Technologies Group, we specialize in maintenance and repair of the QPix 420 Microbial Colony Picking System from Molecular Devices. By concentrating on this high-precision platform and keeping a comprehensive stock of spare parts, we can provide fast, reliable service that keeps your lab running smoothly.

High-performance instruments like the QPix 420 require consistent, expert care to ensure precise colony detection and imaging. Even minor disruptions can halt critical workflows. Thousands of researchers trust ATG to keep their systems operating efficiently—you can too. You should too.

Inquire About Service Contracts

Tips for keeping your Molecular Devices™ QPix 420 Microbial Colony Picking System in good working order:

Daily Maintenance:

  • Ensure that the interior of the instrument is free from dirt and dust. Check the surface of the x-drive drag-chain support bracket. Dust and debris collected here can be swept onto the instrument bed during operation.

  • Before storing or cleaning the head, wipe the exterior of the head with a cloth using 70% ethanol. To store the head, slide it into its metal cover with the pins facing inward and then secure it in place with the thumbscrew and washer.

  • Check the air regulator for signs of moisture. If necessary, twist the drain clockwise approximately a half turn to force out moisture. Some types of regulators might require the drain purge to be pushed up towards the regulator bowl.

Weekly Maintenance:

  • If you use the external air compressor, drain the accumulated moisture from the internal air tank weekly.

  • If the instrument door shows signs of damage, a replacement door needs to be installed.

Having technical problems with your Molecular Devices™ QPix 420 Microbial Colony Picking System?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 
Submit Service Request

Molecular Devices™ QPix 420 Microbial Colony Picking System Troubleshooting Tips:

If One or more of the axes do not move:

  • Make sure that the main power switch is turned on at the wall, the Emergency Stop button is pulled out, and the Start button is pressed.

If The drive system fails to home the actuator:

  • Make sure that the door is closed.

  • Manually move the head to the center of the instrument then close the door and retry the process.

The UV light does not turn on:

  • Make sure that the door is closed.

  • Make sure that filaments in the lamp are not burned out.