When your Cytiva™ Xcellerex™ XDR Bioreactor needs servicing, you deserve specialists who can restore its performance without missing a beat. At Analytical Technologies Group, we specialize in maintaining and repairing bioreactors like the XDR series, empowering our field engineers with the advanced know-how to diagnose and fix issues efficiently. Supported by a comprehensive stock of replacement parts, we deliver fast, reliable service that minimizes downtime and keeps your lab operating at full capacity.

Count on us to get your system running seamlessly — and to keep it performing at its best.

Inquire About Service Contracts

Tips for keeping your Cytiva™ Xcellerex™ XDR Bioreactor in good working order:

Bianually:

  • Test alarm stack functionality

  • Test emergency stop functionality

  • Verify agitator alignment

  • Verify agitator speed accuracy

Annually and after system relocation:

  • Inspect cables and connectors

  • Inspect jacking bolts

  • Inspect pressure relief valve

  • Inspect the I/O Cabinet

  • Inspect vent and drain valves

Having technical problems with your Cytiva™ Xcellerex™ XDR Series Bioreactor?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 
Submit Service Request

Cytiva™ Xcellerex™ XDR Series Bioreactor Troubleshooting Tips:

If the agitator drive is not func tioning as expected:

  • The circuit breaker might be tripped; Reset the circuit breaker.

  • The motor might be overloaded.Acknowledge the agitator fault and try to restart the motor.

If the condensate accumulates in the condensate outlet tubing, the condenser bag might be installed incorrectly. Reinstall the condenser bag.

If there is no power supply to the I/O Cabinet:

  • The circuit breaker might be tripped. Verify that the condition which caused the trip is identified and fixed, and reset the circuit braker.

If the filter heater temperature does not increase:

  • The filter heater cable might not be connected to the I/O Cabinet; Connect the filter heater cable to the I/O Cabinet.

  • The wrong filter heater cable might be connected to the I/O Cabinet; Verify that the appropriate filter heater cable is connected to the I/O Cabinet