When your Cytiva™ ReadyToProcess WAVE 25 Bioreactor requires service, you need experts who can get it running smoothly without delay. At Analytical Technologies Group, we focus on the maintenance and repair of bioreactors like the ReadyToProcess WAVE 25, equipping our field engineers with the specialized expertise to quickly identify and resolve any performance issues. With a well-stocked inventory of replacement parts, we ensure rapid turnaround times that minimize downtime and restore your lab’s productivity.
We’ll get your system back on track quickly — and keep it that way.
Tips for keeping your Cytiva™ ReadyToProcess WAVE 25 Bioreactor in good working order:
Calibration is recommended if high precision is required or when running perfusion with auto calibration disabled. To calibrate the pump:
Enter the inner diameter of tubing used.
Set Rate mode to Flow or rpm. Calibration will be performed in terms of flow rate or pump speed according to this setting.
Enter the desired setpoint (as flow rate or rpm according to the Rate mode setting).
Click Settings →Calibrate and choose which pump to calibrate and enter the desired calibration run time, then click Start pump. Run the calibra-
tion for sufficient time to collect a reliably measurable volume of liquid.
When the pump stops, enter the collected volume and click Finish calibra-
tion.
To prevent microbial or cross contamination, ReadyToProcess WAVE 25 should becleaned after each cultivation. The system must be turned off and unplugged before cleaning. Clean as follows:
Clean the exterior of the system units with a damp cloth and a suitable cleaning agent.
Make sure to clean the temperature sensor arms on the underside of the rocker platform. If dirt accumulates in these arms, they may not function properly, which could cause incorrect temperature regulation.
Having technical problems with your Cytiva™ ReadyToProcess WAVE 25 Bioreactor?
Fast Response
We provide very prompt responses to customers within one business day.
Fast Turnaround
Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.
Expert Staff
ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.
Parts In Stock
To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.
Cytiva™ ReadyToProcess WAVE 25 Bioreactor Troubleshooting Tips:
If the rocking speed varies:
This is normal when DO control is enabled. If not required, disable DO control from the process picture.
If the power button flashes red:
Verify the connection between the rocker and the client computer or network.
Disable the missing components in UNICORN.
Connect the missing components to the rocker.
If heating is too fast or too slow:
The bag size might be incorrect; Check and if necessary reset the bag size setting.When changing the bag size, the rocking must be turned off.
The weight measurement might be incorrect; Make sure that the rocker is placed in a horizontal position. Check that the weight shown in the Process Picture matches the actual weight of the culture medium. If not, tare the scale with the weight of the content in the Cellbag entered as Net Weight.