Agilent® TOF LC/MS and Q-TOF LC/MS System Service & Repair
You rely on your equipment. Your equipment should rely on ATG.
Analytical Technologies Group provides specialized maintenance and repair services for Agilent® 6200 series Time Of Flight LC/MS and 6500 series Quadrupole Time Of Flight LC/MS Systems, as well as other Agilent® mass spectrometers. By concentrating on a select group of product lines, our service engineers develop and sustain an exceptional depth of expertise in the Agilent® TOF LC/MS and Q-TOF LC/MS Systems and related instruments. We maintain an extensive inventory of parts to ensure rapid response and minimal downtime whenever service is required. Through our comprehensive service contract plans, your Agilent® TOF LC/MS and Q-TOF LC/MS Systems receive the dedicated support needed to deliver consistent performance and long-term reliability.
Tips for keeping your Agilent® TOF LC/MS and Q-TOF LC/MS System in good working order:
Flush your nebulizer daily or at the end of each shift (or more often) to flush traces of samples and buffers out of the tubing, valves and nebulizer.
Clean the electrospray spray chamber daily or at the end of each shift or anytime you suspect carryover contamination from one sample or analysis to another.
Clean the dual electrospray spray chamber weekly or whenever symptoms indicate that contamination exists in the spray chamber and normal daily cleaning does not correct the problem.
Having technical problems with your Agilent® TOF LC/MS and Q-TOF LC/MS?
Fast Response
We provide very prompt responses to customers within one business day.
Fast Turnaround
Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.
Expert Staff
ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.
Parts In Stock
To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.
Agilent® TOF LC/MS and Q-TOF LC/MS System Troubleshooting Tips:
If the electrospray spray chamber shows signs of contamination:
Prepare the mobile phase you have been using, and turn off the spray chamber.
Remove the electrospray nebulizer, then open the spray chamber.
Rinse the interior of the spray chamber with the current mobile phase or with a mixture of isopropanol and water.
Wipe the interior of the spray chamber with a clean, lint- free cloth, then rinse the area around the spray shield.
Dampen a clean cloth with the mobile phase, wipe the spray shield and the area around the spray shield.
Close the spray chamber.