At Analytical Technologies Group, we specialize in the maintenance and repair of CTC PAL Systems. Because we focus on these platforms and keep a deep inventory of critical spare parts, we can deliver fast, dependable service that keeps your sample preparation and injections running at full capacity.

Versatile autosamplers need consistent, expert care to keep injections precise, object positions accurate and carryover low. A worn needle seal, a grooved valve rotor or a misaligned injection unit can halt your sample throughput and put your results at risk. For years, experienced scientists have trusted ATG to keep their PAL Systems performing exactly as designed. You should too.

Tips for keeping your CTC PAL System in good order:

Before Each Sequence:

  • Prime the syringe before you start; prime gastight syringes by hand.

  • Check that the plunger moves smoothly and the plunger screw is tight.

  • Fill the wash solvent reservoirs with fresh solvent that suits your samples.

  • Empty the waste container.

  • Check that vial caps are crimped tightly and don’t turn by hand.

Weekly:

  • Clean the syringe, the instrument and its work surfaces.

  • Wipe the outside with a soft, lint-free cloth dampened with mild soap and water.

Annually:

  • Replace the injection valve rotor (more often with heavy use or dirty samples).

  • Install the preventive maintenance kit for your setup (PM Kit 1 for LC, PM Kit 2 for GC).

Having technical problems with your CTC PAL System?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 

CTC PAL System Troubleshooting Tips:

Sample backing up at the needle guide (LC):

  • Use the correct needle gauge for the valve.

  • Replace a leaking valve needle seal.

  • Adjust the valve needle penetration depth.

  • Reduce the injection speed.

Poor reproducibility:

  • Increase the pull-up delay to even out syringe pressure.

  • Reduce the sample volume in the vial to avoid creating a vacuum.

  • Use a gastight syringe and tray cooling for volatile solvents.

  • Reduce the fill speed.

  • Re-crimp any caps that turn by hand.

Unexpected peaks:

  • Adjust the needle depth so the syringe doesn’t draw air.

  • Make sure the air volume setting in the method is zero.