Service for Agilent® Single Quad GC Mass Spectrometers


You rely on your equipment. Your equipment should rely on ATG.

Analytical Technologies Group has deep expertise in maintaining and repairing Agilent® Single Quad GC Mass Spectrometers. We keep critical parts on hand and employ only the most knowledgeable field service engineers, so we can diagnose problems quickly and get your GC/MS back to producing reliable data with minimal downtime. With more than 20,000 instruments serviced, ATG has the experience to help your instruments have a long, productive life.

More than 20,000 instruments serviced. Put yours in experienced hands with ATG.

  • Agilent® 5973 Series MSD

  • Agilent® 5975 Series MSD

  • Agilent® 5977 Series MSD

Tips for keeping your Agilent® Single Quad GC Mass Spectrometer in good order:

Weekly Maintenance:

  • Check the foreline pump oil level. It should be above the lower line and almost clear. If it is dark or cloudy, replace it.

  • Review your latest tune report and compare it with previous ones. Rising electron multiplier voltage or falling abundances are early signs that service is needed.

Monthly Maintenance:

  • Leak check the system, paying close attention to the inlet septum, column nuts and any seals that were recently opened.

  • Check the column nuts. New ferrules can loosen after a few heating and cooling cycles.

Every 6 Months:

  • Replace the foreline pump oil (every 3 months for CI systems using ammonia reagent gas). Treat used oil as hazardous waste.

  • Check the calibration vial and refill it if it is low.

Having technical problems with your Agilent® Single Quad GC Mass Spectrometer?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 

Agilent® Single Quad GC Mass Spectrometer Troubleshooting Tips:

No peaks in the chromatogram:

  • Run autotune. If the MSD passes, the problem is most likely in the GC.

  • Check that the syringe is installed correctly, the sample vial is not empty, and the injection was not made in split mode by mistake.

  • Check the inlet for leaks or contamination.

Autotune fails:

  • Check that the calibration vial is not empty.

  • Check that the vacuum readings are normal for your carrier gas flow.

  • Try the other filament.

Poor sensitivity:

  • Confirm that the correct tune file is loaded for your analysis.

  • Check all temperature setpoints (oven, GC/MSD interface, ion source and mass filter).

  • Check the split ratio and splitless purge time.

  • Clean or replace the inlet liner. If sensitivity is still low, the ion source may need cleaning.

ATG: YOUR PARTNER IN BIOPROCESS SERVICE, SUPPORT and MAINTENANCE