Service for Agilent® 7200 Accurate-Mass Q-TOF GC/MS


You rely on your equipment. Your equipment should rely on ATG.

Analytical Technologies Group specializes in the maintenance and repair of the Agilent® 7200 Accurate-Mass Q-TOF GC/MS. We focus on servicing only a narrow range of product lines which allows our service engineers to maintain an unparalleled knowledge of the Agilent® 7200 Accurate-Mass Q-TOF GC/MS and other lines we service. We keep parts in stock to be ready whenever a problem arises, ensuring your downtimes are always minimal. With our service contract plans, your Agilent® 7200 Accurate-Mass Q-TOF GC/MS is supported for lasting reliability and consistent performance throughout its operational life.

Thousands of customers can’t be wrong relying on ATG. You should too.

  • Agilent® 7200 Accurate-Mass Q-TOF GC/MS

Tips for keeping your Agilent® 7200 Accurate-Mass Q-TOF GC/MS in good order:

Weekly Maintenance:

  • Check the color and level of the foreline pump oil. It should be above the lower line and almost clear. If it is dark or cloudy, replace it.

  • Review your latest tune report and compare it with previous ones to catch performance changes early.

Every 6 Months:

  • Replace the foreline pump oil and the oil mist filter, or sooner if needed. Treat used oil as hazardous waste.

  • Check the calibration vial(s) and refill if low.

As Needed:

  • Check the internal reference mass (IRM) vial. The IRM corrects the mass axis during runs, so do not let it run dry.

  • Replace the GC carrier gas traps and gas supplies, including the nitrogen supply for the collision cell.

  • Clean the ion source when performance shows it is needed. Use the removable ion source to take it out without venting the instrument. CI sources need cleaning more often than EI sources and may look clean even when they are not.

Having technical problems with your Agilent® 7200 Accurate-Mass Q-TOF GC/MS?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 

Agilent® 7200 Accurate-Mass Q-TOF GC/MS Troubleshooting Tips:

No peaks in the chromatogram:

  • Run autotune. If the MS passes, the problem is most likely in the GC.

  • Check that the syringe is installed correctly, the sample vial is not empty, and the injection was not made in split mode by mistake.

  • Check the inlet for leaks or contamination.

Results are not reproducible after startup:

  • After the “Okay to run” message appears, wait 2 hours for the instrument to reach thermal equilibrium before acquiring data.

  • After a vent, major service or power outage, run autotune starting from the default settings, not from the previous tune values.

The foreline pump keeps gurgling during pumpdown:

  • The gurgling should stop within about a minute. If it continues, there is a large air leak, most likely at the analyzer sideplate seal or the interface column nut.

  • If the turbo pumps do not reach 80% speed within 10 to 15 minutes, the system will shut down the foreline pump. Fix the leak, then power cycle the instrument.

ATG: YOUR PARTNER IN BIOPROCESS SERVICE, SUPPORT and MAINTENANCE