When your Agilent® 7000 Series Triple Quadrupole GC/MS requires service, you need experts who can get it running smoothly without delay. At Analytical Technologies Group, we focus on the maintenance and repair of GC/MS systems like the 7000 series, equipping our field engineers with the specialized expertise to quickly identify and resolve any performance issues. With a well-stocked inventory of replacement parts, we ensure rapid turnaround times that minimize downtime and restore your lab’s productivity.

We’ll get your system back on track quickly — and keep it that way.

  • Agilent® 7000 Series Triple Quadrupole GC/MS

Tips for keeping your Agilent® 7000 Series Triple Quadrupole GC/MS in good order:

Weekly Maintenance:

  • Check the foreline pump oil level. It should be above the lower line and almost clear. If it is dark or cloudy, replace it.

  • Review your latest tune report and compare it with previous ones. Watch for rising electron multiplier voltage or changes in abundance.

  • Check that the foreline and analyzer chamber pressures are normal for your carrier and collision gas flows.

Every 6 Months:

  • Replace the foreline pump oil. Treat used oil as hazardous waste.

  • Check the calibration vial and refill it if it is low. After refilling, always purge the calibration valve before tuning. Skipping the purge can damage the filaments and electron multiplier.

As Needed:

  • Replace GC carrier gas traps and gas supplies, and check that the collision cell gas supply is not running low.

  • Never use hydrogen as a collision cell gas.

  • Clean the ion source when tune reports or data show it is needed, for example rising electron multiplier voltage, poor sensitivity or high background.

Having technical problems with your Agilent® 7000 Series Triple Quadrupole GC/MS?

 

Fast Response

We provide very prompt responses to customers within one business day.

Fast Turnaround

Our field service engineers are expediently dispatched across North America & Europe to fulfill your service request.

Expert Staff

ATG’s highly trained, highly skilled staff ensure your instruments remain in optimal condition.

Parts In Stock

To minimize instrument downtime, we stock virtually all parts for the systems in which we specialize.

 

Agilent® 7000 Series Triple Quadrupole GC/MS Troubleshooting Tips:

No peaks in the chromatogram:

  • Run autotune. If the MS passes, the problem is most likely in the GC.

  • Check that the syringe is installed correctly, the sample vial is not empty, and the injection was not made in split mode by mistake.

  • Check the inlet for leaks or contamination.

No peaks or low signal in MRM mode:

  • Confirm that the correct precursor and product ions are selected in the method.

  • Check that the dwell time is not too short and that the collision cell gas is flowing.

  • Check the collision cell voltages (collision energy) in the method.

Autotune fails or there is not enough signal to begin tune:

  • Check that the calibration vial is not empty.

  • Check that the vacuum readings are normal.

  • Check the collision cell gas flow.